    Sensitive Electrochemical Analysis of Gentamicin Sulfate in Eye Solution

    Ekşi, Osman; İpek, Yeliz; Koçak, Berna

    In this study, for the first time, chromium (III) oxide-glutardialdehyde modified boron-doped diamond electrode (Cr2O3-GA2/BDDE) was used as a sensor for the electrochemical analysis of gentamicin sulfate. The Cr2O3-GA2 suspension used in electrode modification was physically coated on the BDD electrode by a simple drop-casting method. Characterization of Cr2O3-GA2 used in the modification was made by scanning electron microscopy, energy-dispersive spectroscopy (SEM-EDS), and the electrochemical method. The behavior of gentamicin sulfate on Cr2O3-GA2/BDDE was studied by cyclic voltammetry (CV), and its electrochemical analysis was done by differential pulse voltammetry (DPV). Electrochemical analysis of gentamicin sulfate on Cr2O3-GA2/BDDE was carried out in a pH 9 Britton–Robinson (BR) buffer by the DPV method. The linear working range of gentamicin sulfate on Cr2O3-GA2/BDDE was calculated as 5.96 × 10−6−6.54 × 10−5mol L−1, and the correlation coefficient was calculated as 0.997. The limit of detection (LOD) was found to be 2.84 × 10−6. mol L−1. Additionally, Genta, a 3% eye/ear drop solution, was preferred to determine the applicability of gentamicin sulfate in a commercial formulation. As a result, it was shown that the method developed on Cr2O3-GA2/BDDE for the determination of gentamicin sulfate can be applied to commercial drugs.